Leave Your Message

News

New Product · JT8040 Humidity Calibrator

News

News Categories
Featured News

New Product · JT8040 Humidity Calibrator

2023-08-21

Temperature and humidity sensors will drift during use, and this phenomenon is impossible to avoid.For a long time, the calibration of relative humidity measuring probes has been carried out in laboratories. The temperature and humidity measuring instruments need to be sent back to the manufacturer's laboratory or metrology agency for calibration, or users need to purchase expensive humidity calibrators or constant temperature and humidity equipment for self-calibration. For general users, this method is neither economical nor convenient.

JANTYTECH Technology has used its deep thermal testing technology foundation and development experience to launch the JT8040 portable humidity calibrator, which solves the pain points of humidity calibration and fills certain industry gaps, providing a faster, more accurate, economical and portable humidity calibration method. It is an ideal choice for users of humidity sensors, portable humidity test instruments, metrology tests, etc.

This series of calibrators provides salt solutions with different humidity values ​​from 23% to 82% for calibration at different humidity values.

JT8040 Humidity Calibrator
Humidity calibrator can replace different saturated salt solution bottles
Humidity calibrator can replace saturated salt solution to achieve calibration of different humidity valuesThe size of the solution bottle cavity can meet the needs of most temperature and humidity probes. Simply insert the temperature and humidity probe into the saturated salt solution bottle cavity to obtain a stable humidity environment, and the humidity value in the cavity can be displayed in real time on the calibrator's display screen.

news1 (1).jpg

Saturated salt solution can be used independently
The saturated salt solution can also be used independently, and the humidity value at the temperature of the measurement can be checked through the standard humidity meter on the bottle. Usually it takes 60 minutes or longer for the humidity to stabilize, depending on the environmental conditions during the measurement.

news1 (2).jpg

Saturated salt solutions available at various humidity levels
Different salt solutions will produce different relative humidity values. JT8040 provides no less than 6 saturated salt solutions with humidity values ​​to choose from, which can be purchased and used separately.

news1 (3).jpg

JT8040 basic structure
Remove  the protective cover of the saturated salt solution tank, and put the salt tank into the JT8040 host cavity, tighten it clockwise
Loosen  the locking nut, insert the temperature sensor into the locking cap, and then tighten the lock nut to secure it.

JT8040-XX Saturated Salt Solution Technical Specifications

Model Name

Technical indicators

Standard value

23%, 33%, 43%, 59%, 75%, 85% (at 20°C) (one standard value for each salt tank)

Testing cycle

>20min (depending on the response time of the sensor being tested)

interface

Standard 15mm (13-18mm can be equipped with JT8040-MF seal)

Sensor insertion length

Less than 60mm

Reuse

No refill (one-time)

Dimensions

110mm (height) x 38mm (diameter)

Material

Transparent acrylic